All repairs that are performed by ScopeCare will meet the following sterilization parameters. Please note that the Non-Autoclavable scopes can not be autoclaved after a repair, since the design of the scope has not changed. Upgrades to autoclavable scopes are available at a competitive price.

Cleaning:

To ensure an optimal image from your recently repaired scope. ScopeCare recommend the the following scope repair program, invloving cleaning and handling procedures to be performed.

  • After use, clean the scope with the appropriate cleaning solution.

  • Rinse the solution with distilled water

  • With a cotton swab, wipe the lens using 90% pure alcohol.

  • Hold the scope up to a light looking on an angle to see if any debris or a rainbow can be seen on the lens. If there is any thing then repeat step III.

  • Prepare the scope for sterilization.

Sterilization

The SCOPECARE scope has been designed to be sterilized through steam autoclave. The following instructions should be followed for sterilization:

  • Wrap the scope in a surgical towel and place in a clean tray.

  • Sterilize at 2.2 bar 273 degrees F (134 C) for three to five minutes.

  • Do not handle a hot scope.

  • Allow the scope to cool in the ambient air. Do not force to cool a scope with a damp towel.

  • If using a SCOPECARE scope sterilization tray. Please insert the scope appropriately into the tray.

ScopeCare scopes have been approved for Steris® and Sterrad® sterilization